Comparing Five Generations of ActiGraph Devices using an Orbital Shaker.
Medicine & Science in Sports & Exercise ( IF 4.1 ) Pub Date : 2025-01-16 , DOI: 10.1249/mss.0000000000003652
Samuel R LaMunion,Joe Nguyen,Robert J Brychta,Richard P Troiano,Karl E Friedl,Kong Y Chen
Medicine & Science in Sports & Exercise ( IF 4.1 ) Pub Date : 2025-01-16 , DOI: 10.1249/mss.0000000000003652
Samuel R LaMunion,Joe Nguyen,Robert J Brychta,Richard P Troiano,Karl E Friedl,Kong Y Chen
INTRODUCTION
ActiGraph accelerometers are used extensively to objectively assess physical activity, sedentary behavior, and sleep. Here, we present an objective validation of five generations of ActiGraph sensors to characterize potential differences in output arising from changes to hardware or firmware.
METHODS
An orbital shaker generated accelerations from 0 to 3700 milli-g in a randomized order to test the wGT3X-BT, GT9X, CentrePoint Insight Watch (CPIW) 1.0 and 2.0, and the GT3X+ devices used in the 2011-2014 National Health and Nutrition Examination Survey (NHANES). Absolute (±50 milli-g) and relative (±5%) raw vector magnitude (VM) agreement to a NIST-calibrated accelerometer (3DM) was the primary outcome for static offset (delta from 3DM VM at 0 g), drift (delta VM start - minus end), and dynamic response.
RESULTS
We observed inter-generation differences [(Mean ± SD) -27.5 ± 8.7 mg (GT9X) to 4.8 ± 32.0 mg (GT3X+)] and intra-generational variability in static offset that carried over to lower shaker accelerations (< ~1300 milli-g), but these variations were mitigated with post-hoc calibration or integrated units (e.g., counts, MIMS). All generations fell within the ±5% of the 3DM across the tested acceleration range and were within ±50 mg of the 3DM below 3000 milli-g's. Drift was small [0.91 ± 0.27 mg (GT3X+), 0.07 ± 0.31 mg (GT9X)] relative to ≥1000 milli-g VM signal.
CONCLUSIONS
Using an orbital shaker and criterion accelerometer we found small differences in the raw output of five ActiGraph generations across a range of accelerations typical of free-living data. However, the devices largely fell within established limits of agreement to the NHANES GT3X+ devices and variability was minimized with auto-calibration or when data was expressed in MIMS or counts.
中文翻译:
使用 Orbital Shaker 比较五代 ActiGraph 设备。
引言 ActiGraph 加速度计广泛用于客观评估身体活动、久坐行为和睡眠。在这里,我们提出了对五代 ActiGraph 传感器的客观验证,以表征因硬件或固件更改而导致的输出潜在差异。方法 轨道振动器以随机顺序产生从 0 到 3700 milli-g 的加速度,以测试 wGT3X-BT、GT9X、CentrePoint Insight Watch (CPIW) 1.0 和 2.0,以及 2011-2014 年全国健康和营养检查调查 (NHANES) 中使用的 GT3X+ 设备。与 NIST 校准的加速度计 (3DM) 的绝对 (±50 milli-g) 和相对 (±5%) 原始矢量幅度 (VM) 一致性是静态偏移(0 g 时 3DM VM 的增量)、漂移(增量 VM 开始 - 负结束)和动态响应的主要结果。结果我们观察到代际差异 [(均值 ± SD) -27.5 ± 8.7 mg (GT9X) 至 4.8 ± 32.0 mg (GT3X+)] 和静态偏移的代内变异性,这些变化延续到较低的摇床加速度 (< ~1300 milli-g),但这些变化通过事后校准或积分单位(例如,计数、MIMS)得到缓解。在测试的加速范围内,所有世代都在 3DM 的 ±5% 范围内,并且在低于 3000 毫克的 3000 毫克的 3DM 的 ±50 毫克范围内。相对于 ≥1000 milli-g VM 信号,漂移很小 [0.91 ± 0.27 mg (GT3X+),0.07 ± 0.31 mg (GT9X)]。结论 使用轨道激振器和标准加速度计,我们发现在自由生活数据的典型加速度范围内,五代 ActiGraph 的原始输出存在微小差异。然而,这些设备在很大程度上落在与 NHANES GT3X+ 设备的既定协议范围内,并且通过自动校准或当数据以 MIMS 或计数表示时,可变性被最小化。
更新日期:2025-01-16
中文翻译:
![](https://scdn.x-mol.com/jcss/images/paperTranslation.png)
使用 Orbital Shaker 比较五代 ActiGraph 设备。
引言 ActiGraph 加速度计广泛用于客观评估身体活动、久坐行为和睡眠。在这里,我们提出了对五代 ActiGraph 传感器的客观验证,以表征因硬件或固件更改而导致的输出潜在差异。方法 轨道振动器以随机顺序产生从 0 到 3700 milli-g 的加速度,以测试 wGT3X-BT、GT9X、CentrePoint Insight Watch (CPIW) 1.0 和 2.0,以及 2011-2014 年全国健康和营养检查调查 (NHANES) 中使用的 GT3X+ 设备。与 NIST 校准的加速度计 (3DM) 的绝对 (±50 milli-g) 和相对 (±5%) 原始矢量幅度 (VM) 一致性是静态偏移(0 g 时 3DM VM 的增量)、漂移(增量 VM 开始 - 负结束)和动态响应的主要结果。结果我们观察到代际差异 [(均值 ± SD) -27.5 ± 8.7 mg (GT9X) 至 4.8 ± 32.0 mg (GT3X+)] 和静态偏移的代内变异性,这些变化延续到较低的摇床加速度 (< ~1300 milli-g),但这些变化通过事后校准或积分单位(例如,计数、MIMS)得到缓解。在测试的加速范围内,所有世代都在 3DM 的 ±5% 范围内,并且在低于 3000 毫克的 3000 毫克的 3DM 的 ±50 毫克范围内。相对于 ≥1000 milli-g VM 信号,漂移很小 [0.91 ± 0.27 mg (GT3X+),0.07 ± 0.31 mg (GT9X)]。结论 使用轨道激振器和标准加速度计,我们发现在自由生活数据的典型加速度范围内,五代 ActiGraph 的原始输出存在微小差异。然而,这些设备在很大程度上落在与 NHANES GT3X+ 设备的既定协议范围内,并且通过自动校准或当数据以 MIMS 或计数表示时,可变性被最小化。