Monte Carlo source verification of Elekta Synergy for pMLC collimated electron beams
Radiation Physics and Chemistry ( IF 2.8 ) Pub Date : 2024-10-31 , DOI: 10.1016/j.radphyschem.2024.112359 D. van Eeden, F.C.P. du Plessis
Radiation Physics and Chemistry ( IF 2.8 ) Pub Date : 2024-10-31 , DOI: 10.1016/j.radphyschem.2024.112359 D. van Eeden, F.C.P. du Plessis
In previous works, an electron beam source was developed that models its energy spectrum with a Lévy distribution. This source is used in BEAMnrc Monte Carlo simulation. It was benchmarked against water tank measurements for electrons delivered through an applicator collimation system, and it was found to agree with 2%/2 mm. This study uses this electron source further by investigating its ability to simulate electrons collimated by the photon multi-leaf collimators for MERT applications. Simulated and Gafchromic EBT3 depth dose and profile data were compared.